Advance Optima series with a modular system design. Broadest range of analysis technology from specialized process photometers to in-line laser measurement. Centrally and remotely operable with Ethernet networks. Designed for use in hazardous areas and with flammable sample gases.
One central unit and up to 4 analyzer modules
– with common control
– with a common connection technology
– in a common system housing
All analyzers use high performance measurement technology
– Photometers
– In-Situ Laser Analyzer
– Oxygen Analyzer with new microwing technology
– Oxygen Analyzers
− Trace Oxygen Analyzer
– Thermal Conductivity Analyzers
– FID Analyzers
Multiple analyzer systems
– up to four analyzer modules
– with up to six sample components
Extensive automatic calibration
– with air or integral calibration cells
– no need for test gas cylinders
Self-monitoring function indicates when maintenance is required
Hazardous areas approved models in accordance with international standards like
– ATEX
– EAC
– CSA
Hazardous area version for FID, AO2040-Fidas24 Ex
– for measuring flammable gases in Zone 1, 2, 21, 22
“Safety Concept” for measuring flammable gases in Zone 2
Provides Ethernet TCP/IP interfaces, OPC server, PROFIBUS and Modbus interfaces
Advance Optima Series - Design Variations | ||
AO2020 | AO2040 | AO2040-Fidas Ex |
19" rack mounted version for controller, display/keypad, power supply and one analyzer module. Additional housings for external analyzer modules possible | Wall mounted version Additional housings for external analyzer modules possible | Special wall mounted version for the hazardous area installation of the FID analyzer (stand-alone). |
Advance Optima Series - General Data |
Size AO2020 | Size AO2040 | Size AO2040-Fidas24 Ex |
483mm W x 412(597)*mm D x 177mm H *values with connection box in brackets | 444mm W x 199mm D x 412(597)* H *values with connection box in brackets | 444 (567)*mm W x 199mm D x 412(601)*mm H *values with connection box and purge controller in brackets |
Housing protection | Housing protection | Housing protection |
IP 20 IP 54 with connection box IP 65 without power supply/display/control unit | IP 20 IP 54 with connection box IP 65 without power supply/display/control unit | IP 65 |
Power supply | Power supply | Power supply |
100…240 VAC, 2,2 A max | 100…240 VAC, 2,2 A max | 110 … 230 VAC, 2 A max. |
The AO2000 series provides the widest range of analyzer modules. Up to four modules can be combined into tailor-made solutions.
All analyzers use high-performance measurement technology and simplified calibration that doesn't require test gas bottles.
NDIR and UV Photometers
In-Situ Laser Module
Oxygen Analyzers
Trace Oxygen Analyzer
Thermal Conductivity Analyzers
FID Analyzers
Type | Name | Sample Gas (Examples) | Application | |
Photometers | ||||
Infrared Photometer | Uras26 | CO, CO2, NO, SO2, N2O, CH4, C3H8, C2H4, ... (max. 4) | Emission, firing, landfill, gas production, burners, fermenters, process monitoring | |
Process Photometer | Limas21 UV | NO, SO2, NO2, H2S, Cl2, COS, ... (max. 3) | Production and processing of Cl2, cellulose, viscose rayon, plastic prod. | |
Process Photometer | Limas21 HW | NO, NO2, NH3, ... (max. 3) | Diesel engine, managing DeNOx, catalyzer | |
In-Situ Laser | ||||
TDLS | LS25 | O2, CO, CH4, NH3, H2O, CO2, HCL, H2S, HCN, HF, ... (max. 2) | Combustion optimization, DeNOx, process measurement and -optimization e.g. in energy-, metal-, chemical-, petrochemical-, glass-, cement- and paper-industry | |
Oxygen Analyzer | ||||
Paramagnetic | Magnos28 | O2 | CEMs, Air fractionating, biochemical plants, purity, with microwing technology | |
Thermo- magnetic | Magnos27 | O2 | Flue gas analysis, roasting gas, cement flue gas | |
Thermal Conductivity | ||||
Silicon sensor | Caldos27 | Ar in O2, H2 in Ar, H2 in N2/Air, ... | Hydrogen purity, turbo generator, inert gas monitoring, UEL/LEL monitoring | |
Glass cell | Caldos25 | H2 in Cl2, SO2 in N2/Air, H2 in N2/Air, ... | Chlorine production, sulfur dioxide in smelter-off gas, ammonia dissociation | |
Flame Ionization | ||||
FID Detector | Fidas24 | Hydrocarbons | Volatile organic compounds, emission monitoring, automotive industries, ambient air, process control |